The international news resource for industrial & municipal water professionals
By Zoe Grosser, Elaine Le Moine January 1, 2007
Trace amounts of personal care products, antibiotics, and birth control hormones have been found in today's wastewater. While the U.S. Environmental Protection Agency (EPA) has developed guidance to help laboratories determine chronic and acute levels of wastewater pollutants, these permissible levels have focused primarily on conventional pollutants. How are laboratory technicians supposed to measure these additional compounds before treatment or after a remediation process has been put into place?
The answer seems to be by using familiar tools together and in new ways.
Gas chromatography (GC), one of the routine “workhorse” techniques in the environmental laboratory, allows specific detectors to be matched to analysis requirements. The combined technologies of GC and a mass spectrometry detector (GC/MS) is the most prominent technique used in the environmental laboratory today because
Monitoring Modes
Still more advances in GC/MS technology have improved detection limits in complex samples. The ability to adjust sensitivity requirements for particular analytical needs makes the system useful for a wider range of applications. Full scan (FS) ion monitoring is by far the most prevalent mode of operation for organic compound detection. It covers the necessary mass range and provides classical spectra that can be compared with those stored in a library for positive identification.
Selected ion monitoring (SIM) is a GC/MS mode of operation in which only the ions of interest are monitored without surrounding interferences and coelutions, providing dramatic increases in sensitivity. SIM has not been widely accepted in environmental testing, because it requires prior knowledge of the sample matrix.
Testing the Instruments
The EPA method for semivolatile organic chemicals in drinking water (Method 525.2) was developed in 1988 using technology that was, at that time, state of the art. The concentration range specified for measurement in the method ranges from 0.1-10.0 nanograms per microliter (ng/µL). Cheri Coody et al. have demonstrated compliance with the requirements of Method 525.2, such as method and sample quality control objectives, for drinking water. The following analysis conditions (Table 1) show the GC/MS parameters required to measure drinking water at the enhanced sensitivity. This work was done by E.A. LeMoine. The system is optimized for the lowest detection limits and calibrated from 0.02-1.0 ng/µL.
Table 1. Gas Chromatography/Mass Spectrometry Conditions for Drinking Water Analysis |
|
|---|---|
Instrument |
These conditions can be used with a PerkinElmer® Clarus® 600 T, Clarus 600 C, Clarus 500, or TurboMass™ Gold GC/MS |
Column |
PE-5MS 30 m x 0.25 mm; 0.25 micrometer film thickness |
Oven temperature program |
40°C for 1 min, 45°C/min to 160°C for 3 min, 6°C to 320°C for 2 min |
Programmable pneumatic control |
Helium @ 1.0 mL/min Injector |
Injector |
250°C |
Injection volume |
1 µL |
Mass scan range |
45-450 mass units (u) |
Scan speed |
2.0 scans/sec |
Filament delay |
3 min |
Ion source temperature |
150°C |
Transfer line temperature |
250°C |
Ionization mode |
Electron ionization |
EPA has developed the solid waste methods for more complicated matrices that might be encountered during hazardous waste cleanup or monitoring. LeMoine et al. evaluated Method 8270, an EPA method for measuring semi-volatile compounds in hazardous waste. Identical standards were analyzed using two sets of experimental conditions. A 50-µL large volume injection was used in both cases. One set of standards was analyzed using the GC/MS FS-50 and the second using SIM-50. Table 2 lists the chromatographic and mass spectrometer conditions used for both experiments.
Table 2. GC/MS Conditions for Solid Waste Matrices |
||
|---|---|---|
Instrument |
These conditions can be used with a PerkinElmer® Clarus® 600 T, Clarus 600 C, Clarus 500, or TurboMass™ Gold GC/MS |
|
Column |
PE --5MS 30 m x 0.25 mm; 0.25 micrometer film thickness |
|
Pre-column |
1 m x 0.32 mm deactivated fused silica |
|
Oven temperature program |
55°C for 5 min, 45°C/min to 160°C, 6°C to 320°C for two minutes |
|
Programmable pneumatic control |
Helium @ 1.0 mL/min |
|
Programmable split/splitless (PSS) injector |
55°C for 4 min; ballistic to 250°C; solvent purge mode |
|
Injection volume |
50 µL |
|
MS mode |
Full scan |
Selected ion monitoring |
Mass scan range |
50 -350 u |
159, 231, 233 u |
Scan speed |
2.0 scans/sec |
2.0 scans/sec |
Filament delay |
5 min |
5 min |
Ion source temperature |
150°C |
150°C |
Transfer line temperature |
250°C |
250°C |
Ionization mode |
Electron ionization |
Electron ionization |
Enhanced sensitivity using the semivolatile drinking water method is shown with a calibration curve for one component (Figure 1). The practical quantitation limits are compared to the maximum contaminant levels (MCL) in drinking water (Table 3). The quantitation limits are well below the level where a decision would be required. In fact, the levels are low enough that, for the majority of compounds, a tenfold reduction in the MCL could be reported reliably.

Figure 1. GC/MS Calibration Curve of Lindane
Table 3. Comparison of the Practical Quantitation Level (PQL) |
||
|---|---|---|
Compound |
PQL (µg/L) |
MCL (µg/L) |
Alachlor |
0.01 |
2 |
Atrazine |
0.01 |
3 |
Endrin |
0.01 |
2 |
Lindane |
0.01 |
0.2 |
Methoxychlor |
0.01 |
40 |
Simazine |
0.01 |
4 |
Calculation based on a 1-L separatory funnel extraction |
||
Meeting the Matrix Challenge
The challenge with a more complicated matrix is to maintain detection limits with the introduction of additional potentially interfering components. Newer MS technology has evolved to better address these considerations. For example, one advance incorporates a prefilter to remove unwanted material before it proceeds to the analytical quadrupole. In addition, the detector is sealed to prevent contamination, making the system more rugged for dirty samples and requiring less frequent cleanup or maintenance.
Large volume injections can increase sensitivity, but the increased solvent can degrade chromatography and hamper system cleanout. Solvent purging (before the analytes of interest are introduced into the column) can provide an efficient analysis of a larger volume sample, allowing less preconcentration and providing enhanced sensitivity. Figure 2 shows how solvent purging works. The solvent purging mode was used with the solid waste extracts, while injecting 50 µL of sample.

Figure 2. Solvent Purge Process for Large Volume Injections
Figure 3 shows the pattern recognition possible and selected ions for toxaphene. Selected ion monitoring is a demonstrated technique for increasing sensitivity, while a full scan is useful for pattern recognition to confirm analyte identification. The use of both techniques on the same chromatographic injection provides primary and supporting information for an accurate identification at lower concentrations. The detection limits obtained in each mode are compared in Table 4.

Figure 3. Pattern and Selected Ions for Toxaphene
Table 4. Detection Limit Comparison |
||
|---|---|---|
| Calibration Peaks | Calculated Analytical Detection Limits | |
| FS-50 (ng/µL) | SIR-50 (ng/µL) | |
| Peak 1 | 0.073 | .0.065 |
| Peak 2 | 0.089 | 0.009 |
| Peak 3 | 0.105 | 0.021 |
| Peak 4 | 0.035 | 0.014 |
| Toxaphene (avg.) | 0.076 | 0.027 |
| Toxaphene MCL: 3 micrograms per liter. Detection limits equivalent to ppb (µL/L) for a 1-L sample concentrated to 1 mL. | ||
When endocrine-disrupting compounds are combined with traditional pollutants, other aspects of analysis become more difficult. One consideration for some of the higher-molecular weight compounds, such as brominated flame retardants, is the need to calibrate with a high molecular-weight material to provide more accurate results. W.D. Goodman et al. and K. Usukura et al. provided a mechanism to calibrate with high molecular weight compounds. A typical flame-retardant analysis will produce ions to 970 mass units (u). To achieve accurate mass calibration in the upper half of this mass range (600-1200 u), a technician must use an alternate mass calibration standard with known ions greater than 600 u. This high mass calibration technique uses a GC injection of the calibration standard triazine, which produces ions up to 1,185 u, allowing calibration of the full mass range.
Wastewater's suite of analytes can vary in number and content and the detection limit requirements can be difficult to achieve. Large volume injection and enhanced spectrometer performance can be used to enhance method sensitivity. Coupled with modern techniques to handle complex matrices in the spectrometer, wastewater analysis becomes more productive when using GC/MS capability and flexibility.
References
1. Coody, Cheri, Michael J. Burke, and Elaine A. LeMoine. 2000. “U.S. EPA Volatile Organics Method 524.2 Using Purge and Trap GC/MS,” PerkinElmer Application Note D-6224.
2. LeMoine, E.A. 1998. “Semivolatile Organics in Drinking Water using EPA Method 525.2 Optimized for Greater Sensitivity,” PerkinElmer Application Note D-5886.
3. LeMoine, E.A. and H. Hoberecht. 2000. “U.S. EPA Method 8270 for Multicomponent Analyte Determination,” PerkinElmer Application Note D-6346.
4. Goodman, W.D. and H. Grecsek. 2006. “Analysis of the High Molecular Weight Flame Retardants - Polybrominated Diphenyl Ethers (PBDEs) by GC/MS,” Poster at Pittcon Conference.
5. Usukura, K., T. Seko, and N. Onda. 2001. “Quadrupole GC/MS Analysis of Polybromodiphenylethers Ranging to m/z of 959,” Analytical Sciences, (17) Supplement, p.579.
This article originally appeared in the January/February 2007 issue of Water & Wastewater Products.
About the author
Zoe Grosser
Zoe Grosser is a Segment Marketing Manager at PerkinElmer Life and Analytical Sciences in Shelton, Conn. She can be reached at 203-402-5320.
About the author
Elaine Le Moine
Elaine Le Moine is a Project Manager at PerkinElmer Life and Analytical Sciences in Shelton, Conn.
American Rivers, a river conservation organization, recently commended the Georgia Environmental Facilities Authority for promoting water efficiency as the first source of supply in its recently released study,